New methods for robust continuous wave T<sub>1ρ</sub> relaxation preparation
Pala, Swetha; Hänninen, Nina E.; Nykänen, Olli; Liimatainen, Timo; Nissi, Mikko J. (2023-01-15)
Pala, S, Hänninen, NE, Nykänen, O, Liimatainen, T, Nissi, MJ. New methods for robust continuous wave T1ρ relaxation preparation. NMR in Biomedicine. 2023; 36( 2):e4834. doi:10.1002/nbm.4834
© 2022 The Authors. NMR in Biomedicine published by John Wiley & Sons Ltd. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
https://rightsstatements.org/vocab/InC/1.0/
https://urn.fi/URN:NBN:fi-fe202301193736
Tiivistelmä
Abstract
Measurement of the longitudinal relaxation time in the rotating frame of reference (T₁ₚ) is sensitive to the fidelity of the main imaging magnetic field (B₀) and that of the RF pulse (B₁). The purpose of this study was to introduce methods for producing continuous wave (CW) T₁ₚ contrast with improved robustness against field inhomogeneities and to compare the sensitivities of several existing and the novel T₁ₚ contrast generation methods with the B₀ and B₁ field inhomogeneities. Four hard-pulse and four adiabatic CW-T₁ₚ magnetization preparations were investigated. Bloch simulations and experimental measurements at different spin-lock amplitudes under ideal and non-ideal conditions, as well as theoretical analysis of the hard-pulse preparations, were conducted to assess the sensitivity of the methods to field inhomogeneities, at low (ω₁ << ΔB₀) and high (ω₁ >> ΔB₀) spin-locking field strengths. In simulations, previously reported single-refocus and new triple-refocus hard-pulse and double-refocus adiabatic preparation schemes were found to be the most robust. The mean normalized absolute deviation between the experimentally measured relaxation times under ideal and non-ideal conditions was found to be smallest for the refocused preparation schemes and broadly in agreement with the sensitivities observed in simulations. Experimentally, all refocused preparations performed better than those that were non-refocused. The findings promote the use of the previously reported hard-pulse single-refocus ΔB0 and B₁ insensitive T₁ₚ as a robust method with minimal RF energy deposition. The double-refocus adiabatic B₁ insensitive rotation-4 CW-T₁ₚ preparation offers further improved insensitivity to field variations, but because of the extra RF deposition, may be preferred for ex vivo applications.
Kokoelmat
- Avoin saatavuus [31995]