Sensitive and quantitative detection of cardiac troponin I with upconverting nanoparticle lateral flow test with minimized interference
Bayoumy, Sherif; Martiskainen, Iida; Heikkilä, Taina; Rautanen, Carita; Hedberg, Pirjo; Hyytiä, Heidi; Wittfooth, Saara; Pettersson, Kim (2021-09-21)
Bayoumy, S., Martiskainen, I., Heikkilä, T. et al. Sensitive and quantitative detection of cardiac troponin I with upconverting nanoparticle lateral flow test with minimized interference. Sci Rep 11, 18698 (2021). https://doi.org/10.1038/s41598-021-98199-y
© The Author(s) 2021. This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
https://creativecommons.org/licenses/by/4.0/
https://urn.fi/URN:NBN:fi-fe2021112356442
Tiivistelmä
Abstract
Measurement of cardiac troponin I (cTnI) should be feasible for point-of-care testing (POCT) to diagnose acute myocardial infarction (AMI). Lateral flow immunoassays (LFIAs) have been long implemented in POCT and clinical settings. However, sensitivity, matrix effect and quantitation in lateral flow immunoassays (LFIAs) have been major limiting factors. The performance of LFIAs can be improved with upconverting nanoparticle (UCNP) reporters. Here we report a new methodological approach to quantify cTnI using UCNP-LFIA technology with minimized plasma interference. The performance of the developed UCNP-LFIA was evaluated using clinical plasma samples (n = 262). The developed UCNP-LFIA was compared to two reference assays, the Siemens Advia Centaur assay and an in-house well-based cTnI assay. By introducing an anti-IgM scrub line and dried EDTA in the LFIA strip, the detection of cTnI in plasma samples was fully recovered. The UCNP-LFIA was able to quantify cTnI concentrations in patient samples within the range of 30–10,000 ng/L. The LoB and LoD of the UCNP-LFIA were 8.4 ng/L and 30 ng/L. The method comparisons showed good correlation (Spearman’s correlation 0.956 and 0.949, p < 0.0001). The developed UCNP-LFIA had LoD suitable for ruling in AMI in patients with elevated cTnI levels and was able to quantify cTnI concentrations in patient samples. The technology has potential to provide simple and rapid assay for POCT in ED setting.
Kokoelmat
- Avoin saatavuus [31657]